Commissioning of two
different algorithms for stereotactic Radiosurgery M6 FI+ CyberKnife system
Mageshraja K.1,
Jefy N.2, Prasobh C.3, Pande S. C.4, Gupta A.5,
Harpreet K.6, Khatri A.7
1Mageshraja Kannan, Medical Physicist, Department of Radiation Oncology, 2Jefy
Ninan, Chief Medical Physicist, Department of Radiation Oncology, 3Prasobh
Chandrasenan, Medical Physicist, 4Subodh Chandra Pande, Director,
Department of Radiation Oncology, 5Aditya Gupta, Director,
Department of Neurosurgery, 6Harpreet Kaur, Medical Physicist,
Department of Radiation Oncology, 7Ajay Khatri, Medical Physicist,
Department of Radiation Oncology; all authors are affiliated with Artemis
Hospital, Sector-51, Gurgaon, Haryana, India
Corresponding
Author: Mageshraja Kannan, Medical
Physicist, Department of Radiation Oncology, Artemis Hospital, Sector-51,
Gurgaon, Haryana, India. Email: mageshraja.k@gmail.com
Abstract
Introduction:
Cyber Knife M6 FI+ a very precise robotic stereotactic radiosurgery system which
is capable of delivering very high radiation dose to the tumour while
minimizing radiation exposure to normal organs and tissues. MultiPlan treatment
planning system is used with the CyberKnife unit and the algorithms used for
optimization are Ray-Tracing and Monte Carlo. Our CyberKnife M6 FI+ system
commissioning is done as per the vendor recommendation and meeting the local
regulatory guidelines. Materials and
Methods: Clinical beam data measurement is carried out using Radiation
Field Analyzer, Diode E, Pinpoint chamber, Semiflex ionization chamber and
Unidose E Electrometer. FC-65 ionization chamber used in absolute dose
calibration. The mechanical accuracy of the robot and image stability was
verified using Radiochromic film (EBT3), E2E and Iris QA toolkits along with
software. StereoPHAN is used for the patient-specific QA point dose
measurement. Results: Tissue phantom
ratio, Off-centre ratio, Output factor, Percentage depth dose, open beam
profile and absolute dose calibration are done as per the protocols. E2E
performed for two different modes- static and motion. Iris aperture size
measured for all the field sizes. The patient-specific QA delivered for both
algorithms. Discussion: Clinical
beam data measurements are within ±1% of composite data set, overall standard
deviation for Output Factors of the fixed and Iris collimators are 0.0026 and
0.0063. The absolute dose was calibrated to 1cGy per MU. E2E, Iris QA and Laser
and radiation coincidence values are within the tolerance. Patient-specific QA
point dose measurement variation for Ray-Tracing and Monte Carlo is 3% and 2%. Conclusion: These exercises are
mandatory to achieve the accurate, precise and high quality of treatment which
also includes patient safety.
Keywords:
Commissioning,
Cyberknife, Monte Carlo, Ray-Tracing, Radiosurgery
Introduction
A CyberKnife M6
FI+ robotic stereotactic radiosurgery system (Accuray, Sunnyvale, CA),
treatment unit in which miniature type linear accelerator has mounted on an
industrial robot [1,2]. CyberKnife with sub-millimetre accuracy it can
treat tumours anywhere in the body like brain, spine, liver, prostate, lung
with the help of frameless real-time image guidance technology and
computer-controlled robotics [1,3]. Due to the high degree of accuracy and
precision, the CyberKnife system is capable of delivering a very high radiation
dose to the target with minimal dose to the nearest critical organs and
surrounding normal tissues [4].
The CyberKnife
linear accelerator emits 6 MV photon with flatting filter free (FFF) at
constant dose rate of 1000 MU/min, 9.3 GHz X Band [3]. The compact lightweight linac
head is attached to a robotic arm that is producing the non-isocentric beam
angles during treatment with 6 degrees of freedom. There are two different types
of secondary collimator systems- Fixed and Iris with 12 different aperture
sizes 5, 7.5, 10, 12.5, 15, 20, 25, 30, 35, 40, 50 and 60mm [3]. Fixed
collimators are having static apertures in size and aperture of Iris collimator
is adjustable under computer control. It contains 2 stacked hexagonal banks of
tungsten segments that together produce a 12 sided aperture a regular dodecagon
[3,5].
Commissioning
beam data are very important to get a good treatment outcome. Because these
measured data consider as a reference and simultaneously used in the MultiPlan
treatment planning system. The quantities required to measure for commissioning
and quality assurance purposes of a CyberKnife system include absolute dose
calculation, using the IAEA TRS-398 protocol [6]. According to Accuray
commissioning recommendation, mechanical accuracy of the robot, and
image stability and patient-specific QA (point dose measurement), Clinical
dosimetry measurements such as tissue-phantom ratios (TPRs), off-center ratio
(OCRs), secondary collimator output factors (OFs), percentage depth dose (PDD)
and open beam profile presented here[3,7]. All the measurement results compared
with Accuray composite beam data set and tolerance values.
Materials
and Methods
All
the clinical beam data (Ray-Tracing and Monte Carlo) acquired using
· Radiation Field Analyzer a computer-controlled
measuring system (SCANLIFT MP3-Therapy beam analyzer: PTW, Freiburg, Germany),
· TM60017 Diode , TM31014 Pinpoint chamber, TM31010
Semiflex ionization chamber (PTW) and
· Unidose E Electrometer (PTW)
The
absolute dose calibration of the accelerator output was accomplished using
· TM30013 Farmer chamber (PTW) and
· Unidose E Electrometer
The
mechanical accuracy of the robot and image stability was verified using
· Radiochromic film (EBT3, Ashland Speciality
Ingredients, Bridgewater, NJ- 08807)
· Film Ball Cube II (EBT2, H.A.Y.E.S. Manufacturing
Services, Sunnyvale, CA-95054)
· Mini Ball Cube II and XLT Phantom kit films (EBT2)
· EPSON Perfection V800 photo scanner
· Head and neck phantom, Synchrony QA tool (Accuray,
Sunnyvale, CA-94089)
· CIRS-Xsight lung tracking phantom kit & 4D
planning phantom (CIRS Tissue Simulation & Phantom Technology, Norfolk,
USA)
· Iris QA software and E2E software (Accuray)
The
patient-specific QA (point dose measurement) was verified using
· StereoPHAN (Sun Nuclear, 3275 Sun tree Blvd,
Melbourne, FL 32940)
· TM31014 Pinpoint chamber
Results
Clinical beam data measurements are performed as per the vendor
recommendations for the commissioning of two different algorithms, Ray-Tracing
and Monte Carlo [3]. The Ray-Tracing algorithm needs TPR, OCR and
OF measurements [3,7].
Monte Carlo algorithm PDD and Open field profiles in addition to the Ray-Tracing
algorithm [3].
A.1.TPR Measurement
The Tissue phantom ratio (TPR) is the ratio of absorbed dose at a given
point to the dose to the dose at a fixed reference depth using constant SAD
[8]. The reference depth for the CyberKnife
system is 15mm (Dmax) for all collimator sizes and SAD is 800mm [3]. The TPR measurement was carried out using
Diode E position in the RFA water phantom.TPR measurement taken in different
depth 0, 3, 5, 8, 10, 13, 20, 30, 50, 100, 150, 200, 250 and 300mm for all the
field sizes for both collimators at constant SAD 800mm (Table 1 and 2). We use
a cubic spline fit to generate a TPR curve depths from 0 to 300mm and normalize
the values for each collimator to the depth of 15mm as shown in Figure 1 and 2.
A.2.OCR Measurement
The OCR at a particular depth is the ratio of absorbed dose at a given
off-axis point relative to the dose at central axis
[8]. OCR measurement carried out using field
chamber of Diode E, reference chamber of the Semiflex ionization chamber and
water phantom. A central check carried out at two different depth 15mm and
200mm for using the 60mm collimator at SSD of 800mm. Normalization is done to
align the radiation beam center to the detector center. Fixed collimator OCR
measured by conducting orthogonal scans across at the depth of 15mm and 100mm.
In Iris collimator same like Fixed collimator scans and additionally rotate the
linac head in 15 degrees and generate orthogonal scans in the same setup.
Because Iris collimator having a Dodecagonal aperture [3].OCR values were calculated by average in
each side of the cross plan and in plan scans. Therefore each entry in the OCR
data table is the average of four measurement values for fixed collimator and
eight measurement values for Iris collimator as shown in Figure 3 and 4.
A.3.Output Factors
The output factor (OF) is the ratio of absorbed dose at a particular
field size relative to the dose at a reference field size
[3,8]. The reference field size for the CyberKnife
system is based on the 60mm fixed collimator at SAD 800mm [2]. The measurement carried out by using the
Diode E, water phantom and Unidose E electrometer. All the measurement carried
out at the depth of 15mm (Dmax). Meter reading is taken five times continually
for each field size and calculated the average value. Both secondary
collimators Fixed and Iris average value of each field size is normalized to
reference field size of 60mm Fixed collimator value[3]. All the measured values are compared with
the composite data set (Accuray) as shown in Table 3.
A.4. PDD Measurement
PDD is defined as the ratio of the absorbed dose at any depth to the
absorbed at a reference depth (Dmax)
[8]. PDD measurements are performed using 60mm
fixed collimator. Check the center of the linac beam at two depths in the water
phantom 100 mm and 200 mm to verify that the linac beam is pointing straight
down. The PDD measurement is acquired at the depth of 1 mm to 300 mm.
A.5. Open Field Profile
Open field profile measurements are done with no collimators attached to
the fixed collimator housing. A diode detector is positioned in the water
phantom at the depth of 25mm from the water surface and SAD 800mm
[3]. Center check is done with PTW MedPhysto
software to make sure the origin of the water phantom is in coincident with the
central axis of the linac radiation beam. A set of orthogonal scan profiles are
acquired extending from -80 mm to +80mm in each direction.
A. The absolute dose calibration of the
accelerator output
Absolute dose calibration of the CyberKnife was accomplished in accord
with the IAEA TRS 398 protocol. The CyberKnife output was calibrated to deliver
1cGy per MU under reference conditions 60mm collimator, 800 mm SAD. Absolute
dose calibration carried using FG 65 ionization chamber, water phantom, and Unidose
E electrometer. To know the KQ,Q0 of the chamber need to find TPR20,10.
TPR20,10 measured at two different depth 200 mm and 100 mm under the
reference condition. The output measured at the reference depth of 100mm
[6].
The Output formalism is
DQ, 100 mm= MR x ND,W x KT,P x KPol
x Kion x KQ,Q0
where DQ =DQ,100 mm/ TPR (100 mm), DQ,100 mm
= Absolute dose at 100 mm depth, DQ= Absolute dose at Dmax, MR = Electrometer
Reading (nC), ND,W= Chamber Calibration Factor (Gy/C), KT,P=Chamber
temperature and pressure correction factor, KPol = Chamber polarity
correction factor, Kion = ion recombination factor, KQ,Q0
=Beam Quality Index and TPR (100mm) = Tissue phantom ratio at 100mm depth
C.
The mechanical accuracy of the robot and image stability
C.1. End-to-End test
The E2E test is used to determine the total positional error for each
stationary tracking mode and motion tracking mode installed on a CyberKnife system
[3]. Stationary tracking modes include the 6D
skull tracking system, the fiducial tracking system, and the Xsight spine
tracking system. Head and Neck phantom and synchrony QA tool are used for
stationary tracking mode. Motion tracking mode includes the Xsight lung
tracking system and Synchrony Respiratory Tracking System. CIRS- Xsight lung
tracking phantom kit is used in motion tracking system [9]. There are two orthogonal Radiochromic films
loaded in phantoms. The plans are generated according to Accuray recommendation
[3] (Table 4). Both tracking
modes plans are delivered with couch positional accuracy of less than 1mm and
10. EPSON Perfection V800 photo scanner was calibrated to scan the two exposed
films (Axial and Sagittal) and one unexposed film. An unexposed film used to
subtract the background during analyzing processes. E2E software is used to analyze
the scanned film data (Table 5). The specification for total positional error
for the E2E test is ≤ 0.95 mm for all stationary and motion tracking modes and
our results were well within the prescribed limits. [3].
C.2. Laser and radiation coincidence test
The CyberKnife system uses a pinhole laser that is coincident with the
radiation field central axis. The LINAC laser is reflected off an adjustable
mirror and aligned to the mechanical center of the collimators. Laser and
radiation coincidence test carried out using 35mm fixed collimator and two
radiochromic films at two different distance. The first film exposed 800MU at
SDD 800 mm the laser point marked in film, the same procedure repeated for the
second film also only change in the SDD 1600 mm. EPSON Perfection V800 photo
scanner was used to scan the exposed film and ImageJ software used to analyze
the scanned data (Table 6). The tolerance value is < 1mm at SDD 800mm and
< 2mm at SDD 1600mm [3].
C.2. Iris QA
Iris is one of the secondary collimators in CyberKnife system and its
aperture size changes are computer controlled. To verify the aperture size in
Iris collimator radiochromic film, Iris QA, Iris QA hardware accessories are used
[3,10]. Birdcage assembly attached to collimator
system and Iris QA film mount placed on birdcage [3,5]. Radiochromic film position on film mount,
Build up 15 mm kept the top of the film. Irradiate the film with 600 MU. Each
aperture sizes are repeated three times. The same step up needs to do for 15 mm
fixed collimator for Iris QA analysis purpose. The irradiated films and blank
films are scanned using EPSON Perfection V800 photo scanner. The Iris QA
software used to analyze the scanned data and is as shown in (Table 7). The
tolerance should be less than ± 2mm of baseline values [3,5].
D. The patient specific QA (point dose
measurement)
The patient specification QA has done for both algorithms Ray-Tracing and
Monte Carlo using StereoPHAN, Pinpoint chamber and Unidose E Electrometer. The
QA plan generated in Multiplan MD suite version 5.3.0 and noted the planned
point dose value for both algorithms. StereoPHAN and pinpoint chamber set in
the treatment position and deliver the 2000MU for warm up. After the zeroing QA
plans are delivered. Electrometer reading, temperature, and pressure values are
noted. Measured Point dose values are found using the absolute dose formalism.
The percentage of variation between planned value and the measured value should
be less than is ±5% [3]. The measurement setup is as shown in Figure 5.
Table-1: TPR value for all fixed
collimators and normalized value to 15 mm depth.
Depth
(mm) |
Collimators (mm) |
||||||||||
5 |
7.5 |
10 |
12.5 |
15 |
20 |
25 |
30 |
35 |
40 |
60 |
|
0 |
0.566 |
0.51 |
0.487 |
0.474 |
0.463 |
0.46 |
0.459 |
0.46 |
0.462 |
0.462 |
0.483 |
3 |
0.824 |
0.773 |
0.717 |
0.706 |
0.7 |
0.678 |
0.678 |
0.683 |
0.681 |
0.693 |
0.695 |
5 |
0.94 |
0.897 |
0.856 |
0.843 |
0.835 |
0.818 |
0.813 |
0.815 |
0.811 |
0.816 |
0.82 |
8 |
1.005 |
0.981 |
0.956 |
0.948 |
0.937 |
0.924 |
0.919 |
0.922 |
0.92 |
0.923 |
0.919 |
10 |
1.013 |
0.999 |
0.985 |
0.977 |
0.969 |
0.968 |
0.959 |
0.96 |
0.959 |
0.957 |
0.959 |
13 |
1.008 |
1.005 |
0.999 |
1 |
0.998 |
0.992 |
0.991 |
0.989 |
0.989 |
0.989 |
0.989 |
15 |
1 |
1 |
1 |
1 |
1 |
1 |
1 |
1 |
1 |
1 |
1 |
20 |
0.982 |
0.986 |
0.986 |
0.987 |
0.99 |
0.998 |
1 |
1.001 |
1.004 |
0.999 |
0.999 |
30 |
0.929 |
0.936 |
0.936 |
0.947 |
0.951 |
0.961 |
0.966 |
0.97 |
0.974 |
0.97 |
0.981 |
50 |
0.832 |
0.843 |
0.847 |
0.86 |
0.859 |
0.878 |
0.884 |
0.888 |
0.895 |
0.898 |
0.909 |
100 |
0.638 |
0.651 |
0.656 |
0.667 |
0.672 |
0.687 |
0.699 |
0.705 |
0.713 |
0.718 |
0.74 |
150 |
0.496 |
0.506 |
0.514 |
0.526 |
0.527 |
0.54 |
0.552 |
0.557 |
0.562 |
0.568 |
0.593 |
200 |
0.382 |
0.399 |
0.405 |
0.414 |
0.418 |
0.429 |
0.439 |
0.442 |
0.45 |
0.456 |
0.475 |
250 |
0.304 |
0.314 |
0.321 |
0.33 |
0.333 |
0.342 |
0.349 |
0.354 |
0.359 |
0.364 |
0.382 |
300 |
0.241 |
0.252 |
0.257 |
0.264 |
0.269 |
0.276 |
0.281 |
0.285 |
0.289 |
0.293 |
0.307 |
Table-2: TPR value for all Iris
apertures and normalized value to 15 mm depth.
Depth
(mm) |
Collimators (mm) |
||||||||||
5 |
7.5 |
10 |
12.5 |
15 |
20 |
25 |
30 |
35 |
40 |
60 |
|
0 |
0.557 |
0.517 |
0.483 |
0.473 |
0.462 |
0.46 |
0.459 |
0.461 |
0.461 |
0.461 |
0.474 |
3 |
0.794 |
0.749 |
0.714 |
0.69 |
0.677 |
0.67 |
0.673 |
0.674 |
0.671 |
0.669 |
0.671 |
5 |
0.925 |
0.884 |
0.859 |
0.828 |
0.822 |
0.81 |
0.813 |
0.8 |
0.811 |
0.809 |
0.81 |
8 |
0.993 |
0.975 |
0.952 |
0.943 |
0.933 |
0.925 |
0.923 |
0.919 |
0.914 |
0.914 |
0.916 |
10 |
1.005 |
1 |
0.985 |
0.976 |
0.968 |
0.96 |
0.965 |
0.961 |
0.959 |
0.955 |
0.96 |
13 |
1.006 |
1.005 |
0.995 |
0.998 |
0.996 |
0.989 |
0.991 |
0.993 |
0.989 |
0.989 |
0.992 |
15 |
1 |
1 |
1 |
1 |
1 |
1.002 |
1 |
1 |
1 |
1 |
1 |
20 |
0.977 |
0.985 |
0.989 |
0.989 |
0.996 |
1 |
1.005 |
1.002 |
1.006 |
0.998 |
1.011 |
30 |
0.931 |
0.934 |
0.938 |
0.946 |
0.953 |
0.965 |
0.973 |
0.97 |
0.971 |
0.975 |
0.98 |
50 |
0.832 |
0.846 |
0.853 |
0.862 |
0.862 |
0.877 |
0.891 |
0.89 |
0.898 |
0.898 |
0.915 |
100 |
0.64 |
0.649 |
0.658 |
0.67 |
0.673 |
0.686 |
0.701 |
0.706 |
0.713 |
0.717 |
0.742 |
150 |
0.495 |
0.507 |
0.517 |
0.525 |
0.533 |
0.544 |
0.551 |
0.557 |
0.565 |
0.565 |
0.596 |
200 |
0.387 |
0.4 |
0.406 |
0.414 |
0.417 |
0.432 |
0.439 |
0.445 |
0.45 |
0.453 |
0.476 |
250 |
0.306 |
0.312 |
0.323 |
0.331 |
0.333 |
0.342 |
0.35 |
0.354 |
0.36 |
0.362 |
0.382 |
300 |
0.24 |
0.251 |
0.26 |
0.265 |
0.269 |
0.275 |
0.282 |
0.287 |
0.287 |
0.293 |
0.307 |
Table-3: Comparison of output factors
with the Accuray provided composite data
Output Factor |
||||||||
Collimator Size |
Fixed Cone |
Iris |
||||||
Output Factor |
Composite |
Ratio To Composite |
Standard Deviation |
Output
Factor |
Composite |
Ratio
To Composite |
Standard
Deviation |
|
5.0 mm |
0.680 |
0.675 |
1.008 |
0.031 |
0.551 |
0.541 |
1.018 |
0.052 |
7.5 mm |
0.834 |
0.829 |
1.006 |
0.024 |
0.805 |
0.796 |
1.012 |
0.026 |
10.0 mm |
0.881 |
0.878 |
1.004 |
0.018 |
0.884 |
0.877 |
1.008 |
0.018 |
12.5 mm |
0.916 |
0.914 |
1.002 |
0.013 |
0.917 |
0.915 |
1.002 |
0.012 |
15.0 mm |
0.939 |
0.938 |
1.001 |
0.008 |
0.939 |
0.938 |
1.000 |
0.009 |
20.0 mm |
0.963 |
0.962 |
1.001 |
0.005 |
0.962 |
0.962 |
0.999 |
0.006 |
25.0 mm |
0.974 |
0.974 |
1.000 |
0.004 |
0.973 |
0.974 |
0.999 |
0.007 |
30.0 mm |
0.981 |
0.980 |
1.001 |
0.004 |
0.980 |
0.980 |
1.000 |
0.005 |
35.0 mm |
0.986 |
0.985 |
1.001 |
0.003 |
0.985 |
0.985 |
0.999 |
0.006 |
40.0 mm |
0.989 |
0.989 |
1.001 |
0.003 |
0.989 |
0.989 |
0.999 |
0.005 |
50.0 mm |
0.995 |
0.995 |
1.000 |
0.003 |
0.995 |
0.995 |
0.999 |
0.005 |
60.0 mm |
1.000 |
1.000 |
1.000 |
0 |
0.999 |
1.000 |
0.999 |
0.005 |
Table-4:E2E planning protocol for both
collimators
Planning
Constraints |
Ball-cube
6D Skull Tracking |
Ball-cube
Fiducial Tracking |
Ball-cube
Synchrony with Fiducial Tracking |
Mini
Ball-cube Xsight Spine Tracking |
Film
Insert (25.4mm ball only) X-sight Lung Tracking with Synchrony |
Anatomy |
Head |
Body |
Body |
Body |
Body |
Field Size |
30 mm |
25 mm |
25 mm |
15 mm |
15 mm |
Dose, cGy |
600 |
600 |
600 |
600 |
600 |
Prescription |
70%- 420cGy |
70%-
420cGy |
70%-
420cGy |
70%-
420cGy |
70%-
420cGy |
Table-5:E2E analysed data values
Tracking
mode |
Fixed
(< 0. 95 mm) |
Iris
(< 0. 95 mm) |
6D
Skull |
0.45
mm |
0.39 mm |
Fiducial |
0.56
mm |
0.54 mm |
Synchrony
with fiducial |
0.30
mm |
0.35 mm |
X
Sight Spine |
0.52
mm |
0.48 mm |
X
Sight Lung |
0.50
mm |
0.45 mm |
Table-6: Laser and radiation coincidence
analysed data values
Collimators |
SDD=800
mm |
SDD=1600
mm |
Fixed |
0.45 |
1.2 |
Iris |
0.51 |
1.15 |
Table-7: Iris apertures exposed film
data measured value
Aperture
size |
Measured
value |
5 mm |
4.80 mm |
7.5 mm |
7.33 mm |
10 mm |
9.84mm |
12.5 mm |
12.38mm |
15 mm |
14.90 mm |
20 mm |
19.91 mm |
25 mm |
24.95 mm |
30 mm |
29.98 mm |
35 mm |
34.90mm |
40 mm |
39.88 mm |
50 mm |
49.90 mm |
60 mm |
59.92mm |
Figure-1:
TPR curve for all fixed collimators and normalized value to 15 mm depth
Figure-2:
TPR curve for all Iris apertures and normalized value to 15 mm depth.
Figure-3:
OCR curves for fixed collimators at 15mm and 100mm depth.
Figure-4:
OCR curves for all Iris apertures at 15mm and 100mmdepth
Discussion
Stereotactic
Radio surgery system is capable of delivering high dose of radiation within 1
mm of the tumour. So the system dose delivery should be very accurate and
precise [1, 4, 11]. Beam data commissioning is the important parameter to
achieve the accuracy and precision of the system. The same data is used to
generate reference data and is also used in the treatment planning system
[1,3,4,12].The collected beam data are of the highest quality to avoid
dosimetric and patient treatment error that may subsequently lead to a poor
radiation outcome[12,13].
The personnel
performing commissioning are recommended to have thorough knowledge about the
algorithm specific beam parameters which are to be measured [4,13,14].
Subhash C [1]
done the commission and acceptant test of CyberKnife. Clinical beam data measurements Tissue phantom ratio,
Off axis ratio and output factor are compared with Accuray multisite data in
USA. They found their results agreeing within ±2% average multisite data. In
our study in addition to TPR, OAR, OF we also measured PDD and open beam
profile and compared with Accuray multisite data. The results were of excellent
agreement. Each of our clinical beam data measurement was within ±1% average
multisite data. A 6MV CyberKnife TPR
data comparison was done by Subhash C [1] with his measurements to that of Day
and Arac [15]. The result was that the difference between the measured TPR
values increase with increase in the measurement depth [1]. In my study we
compared both of their results with our data. The difference in our data with
that of Subhash C[1] followed the same pattern with maximum difference at 30cm
depth. The output factor (OF) for both collimators was compared with composite
data from Accuray (Table 3). Fixed and Iris 5mm collimator has the maximum
standard deviation in output factor and highest output factor ratio to
composite (0.031, 0.0052) & (1.008, 1.018). The overall standard deviation
in output factor of the fixed and Iris collimators are 0.0026 and 0.0063.
Francescon [16]
compared the different detectors used for measuring clinical beam data in M6 CyberKnife system. They noted that in
small aperture sizes there is a variation in off axis ratio, percentage depth
dose and output factor with various detectors. The Air filled chambers due to
average volumetric effect underestimate measurements and semi conductor
detector overestimate measurement [16,17]. In our measurements we noticed that
Diode E detector over estimate the range +5% and pinpoint chamber underestimate
the range -7% for 5mm collimator size.
International
Atomic Energy Agency TRS 398 [6] recommendation for external beam radiotherapy
is to calibrate the linac to deliver 1 cGy is equal to 1 MU at reference
conditions. According to their recommendation the CyberKnife absolute dose was
calibrated to 1cGy per MU and the beam quality index is 0.670 which was within
the tolerance of 6MV photon beam value (0.676±0.009) [6].
AAPM Task Group
no 135 [18] recommends E2E test as one of the major QA for mechanical accuracy
of the CyberKnife system. It is
recommended to perform E2E test at
least for one stationary mode
and one motion tracking mode on a monthly basis. The maximum difference
between the centers of the planned dose and delivered dose must not exceed 0.95
mm for static treatments and 1.5 mm for motion-tracking treatments. The E2E
test for stationary modes are performed with help of head phantom .An
isocentric treatment plan covering the target sphere with the 70% isodose line
is created. This plan is then delivered and a comparison of the position of the
70% isodose line dose distribution with the known centroid position is
performed [3, 18]. The results are shown in the (Table 5). Ideally CyberKnife
system well-calibrated typically performs static E2E tests on the level of
0.3–0.7 mm [18]. Our results for both stationary
and motion tracking modes were well within the tolerance level.
Iris aperture
size verification is among the monthly quality assurance of CyberKnife radio surgery system
[3,18]. The Iris collimator consists of two stacked hexagonal banks of tungsten
segments together produce a 12-sided aperture at nominal distance of 800 mm.
Iris is a computer-controlled collimator which benefits in improved plan
quality and time efficiency [3,5,19]. Several authors recommend different
techniques using ion chamber, optical image based and Radiochromic film to find
out the Iris aperture size. Sarah CH [5] verifies the aperture size using the
pinpoint ionization chamber. They have the base line output factor for all 12
aperture size during the commission time. This method is economically efficient
and less time consuming. Although the detector positional uncertainty is more
for small collimators 5mm and 7.5 mm in this method. But in our study high
resolution Radiochromic film is used to measure the different aperture sizes
[3]. The results are shown in the (Table 7). With Iris aperture we found a
maximum deviation to 4.80 mm for 5 mm collimator, although within the tolerance
limit [3, 18]. Drawback of this method is that it is time consuming. The whole
film analysis process can easily take up to 2 hours.
In stereotactic
radiosurgery even a small error in treatment planning, delivery, or dosimetric
can lead to poor radiation outcome. Before start of the patient treatment
patient specific quality assurance should be performed [8, 20]. Our planning
station have two different algorithms Monte Carlo and Ray-Tracing. Monte Carlo
is more accurate compared to Ray Tracing in lung and other heterogeneous tissue
[21]. Drawback of Monte Carlo is that it takes more time in dose calculation.
In our study plans are generated using both algorithms and implemented in
phantom. The point dose difference between planned and delivered is less than
5%. The values are well within the range given by vendor [3, 8].
Conclusion
All the clinical
beam data are well in agreement, utilized as input to the Monte Carlo and Ray
Tracing algorithm in Multiplan treatment planning system for further clinical
use. The result of mechanical accuracy tests like E2E and Iris QA has shown
good stability of machine. Patient specific QA results gave us more confidence
to deliver high radiation dose to treating patients. These exercises are
mandatory to achieve the accurate, precise and high quality of treatment which
also includes patient safety.
References
How to cite this article?
Mageshraja K., Jefy N., Prasobh C., Pande S. C., Gupta A., Harpreet K., Khatri A. Commissioning of two different algorithms for stereotactic Radiosurgery M6 FI+ CyberKnife system. Int J Med Res Rev 2018; 6(08): 477-486. doi:10.17511/ijmrr.2018.i08.13.